ABSTRACT
PURPOSE:
Evaluation of the effects of a vehicle's movement/ operation on fluid flow regulators during patient transport. To determine whether or not vehicle trembling during idling and movement during patient transport are factors affecting the velocity of fluid injection.
METHODS:
The volume of fluid, as measured in an idling or moving vehicle, was evaluated using three different types of marketed flow regulators at three different injection speeds: 10, 20 and 40 ml/hr.
RESULTS:
In all cases, when the vehicle was idling or in motion, discrepancies were observed between the pre-calculated amount of fluid and the actual amount of fluid injected. However, a greater discrepancy was observed to exist in a moving vehicle.
CONCLUSION:
The trembling and movement of a vehicle during patient transport affects fluid injection results.
Key words: Infusion pump, Transportation of patients, Fluid therapy
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,431
View
7
Download
Related article
Editorial Office
The Korean Society of Emergency Medicine
TEL: +82-62-226-1780   FAX: +82-62-224-3501   E-mail: 0012194@csuh.co.kr
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © The Korean Society of Emergency Medicine.                 Developed in M2PI