A Case of Pacemaker-mediated Tachycardia
Hyo Yong Ahn, Young Gwan Ko
ABSTRACT
Pacemaker-mediated tachycardia(PMT) is a circus movement tachycardia that can occur in patients with dual chamber pacemaker with atrial sensing(VDD or DDD), when a ventricular depolarization, either spontaneous or paced, causes retrograde activation of the atrium. Recently, a number of pacemaker manufactures have incorporated in their devices a variety of relatively complex algorithms to prevent PMT. Despite these measures, PMT may still occur because of inappropriate programming or unpredictable variations of ventriculoatrial conduction. We report one case of PMT in a 78-year-old male who received DDD type pacemaker due to sick sinus syndrome. In this case, PMT was disappeared after reprogramming parameters of pacemaker and the application of PMT protection algorithm.